News · Explainers · 5 min read
A transistor that leaks: how REDMOON counts radiation
Our radiation payload weighs 42 g and uses about a tenth of a watt. Inside are sensors that measure dose by slowly losing charge. Here's how.
Dose in, frequency out · a floating-gate dosimeter, simplified · Drawing: Lunar Zebro
REDMOON, the radiation payload for our first rover, weighs 42 g and uses about 128 milliwatts, less than a bicycle light. Radiation monitors are usually big and power-hungry. So how does something this small measure radiation? Its sensors are transistors that slowly leak.

Why measure radiation on the Moon
On Earth, the atmosphere and the magnetic field shield us. The Moon has neither. Particles from the Sun and from far outside the Solar System hit the ground directly, and some knock new particles loose from the soil, which add up to about a fifth more. The Moon itself does block about half of the particles from deep space, simply by being in the way.
That radiation flips bits in computer memory, slowly wears out electronics, and is a risk for every astronaut who will live and work there. Only one instrument has measured it actively on the surface, on China's Chang'e-4 lander, and it stayed in one spot. A rover could measure it from place to place: in the open, next to rocks, at the edge of a crater.
A crater wall does two things at once. It hides part of the sky, so fewer particles arrive from space. But particles that hit the wall knock new ones loose, and some of those come the rover's way. Which effect wins, and by how much, is exactly the kind of question a walking dosimeter can answer. A lava tube goes further: rock on every side, and sky only straight above. Future bases may shelter in one, and a small rover could measure how well it shields before anyone moves in.
A gate that leaks
REDMOON is built around four floating-gate dosimeters (FGDOS), chips made by Sealicon together with CERN.
Every transistor has a gate: a small electrode that switches the current through it. In a floating-gate transistor, that gate is sealed off in insulation, with no wire attached. You put a charge on it once and it stays there. The same trick keeps the data on a USB stick when it's unplugged.
When radiation passes through the insulation, it frees pairs of electric charges. They drift towards the gate and cancel part of its charge. The more radiation, the less charge is left. So the gate is a tiny tank that drains a little with every particle, and measuring what's left tells you the dose.
Dose in, frequency out
Reading that charge directly is fiddly, so the chip does it for you: it turns the charge into a frequency. As the dose adds up, the frequency falls, and it falls in a straight line. That's what makes the sensor easy to use: one number, its sensitivity in kilohertz per gray, turns a change in frequency into a dose.
A tank that only drains would run dry. So when the frequency reaches the bottom of its straight range, the chip recharges the gate and starts counting again, like the sawtooth in the drawing. That keeps it in its straight range and stretches how much dose it can measure.
Each chip also carries a thermometer and a reference that radiation doesn't affect, so the readings can be corrected for temperature. On the Moon that matters.
Counting while switched off
Here's the clever part for a rover. The gate drains whether the chip is powered or not. While the rover sleeps through the lunar night, or rides to the Moon switched off, its dosimeters keep counting. When it wakes up, it reads how much charge is gone.
That's why REDMOON can start measuring in Earth orbit, keep going during the trip, and then record the dose along the rover's path. The 2022 design puts one board inside the rover's body and a small one outside, under the solar panel, so it can also measure how well the rover's own body shields its electronics.
Small numbers, big range
| REDMOON | |
|---|---|
| Sensors | 4 floating-gate dosimeters |
| Mass | 42 g, including a 20% margin |
| Power | about 128 mW, including a 20% margin |
| Dose it must handle | at least 100 Gy, about what we expect over the whole mission behind 2 mm of aluminium |
| Smallest step it must see | 0.5 mGy |
What the tests taught us
A sensor like this has to be calibrated: the datasheet's sensitivity isn't good enough. Earlier research measured about 30 kHz per gray where the datasheet promised 70. So we took REDMOON's sensors to three very different beams.
At HollandPTC in Delft, William De Meyere and Abhimanyu Shanbhag fired protons at them, also at different temperatures, which hadn't been done before for these sensors. They found a new effect, which they called the "frequency surge". At the Reactor Institute Delft and at CERN's CHARM facility, Stan Broere tested them with gamma rays at the low dose rates of the Moon and with a mixed field of many particles.
The verdict: promising, not yet flight-ready. The sensors respond in a straight line, as they should, but were less sensitive than expected at two of the three facilities, a recharging bug still needs fixing, and the power use rises with dose. The story of those tests is in our post from CERN.
Read: our radiation payload at CERN REDMOON on the Science page All radiation theses